Radiation Exposure to the Personnel Performing Myocardial Blood Flow Quantification Study Using 13N-ammonia Positron
Nivedita Rana1, Ashish Srivastava1, Munish Kumar1
1Department of Nuclear Medicine, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
Purpose:
The present study aimed to evaluate radiation exposure to staff performing coronary flow reserve (CFR) measurement using 13N-ammonia.
Materials And Methods:
The radiation exposure rate during the administration of 13N-ammonia for the rest and stress part of the study was noted using an ionization chamber-based calibrated survey monitor. The radiation exposure to persons involved in dispensing radioactivity (D1), administering radioactivity (D2) and monitoring the patient during pharmacological stress (D3) were measured using an energy compensated Si-diode personal pocket dosimeter.
Results:
The average dose received by individuals with dosimeters D1, D2, and D3 was 1.28 ± 0.79 µSv, 1.56 ± 0.51 µSv, and 0.88 ± 0.97 µSv per injection, respectively, during the rest of study and 1.56 ± 0.96 µSv, 2.64 ± 1.22 µSv, and 2.2 ± 1.7 µSv per injection, respectively, during stress study. The average exposure rate during the administration of 13N-ammonia at 0.5 m and 1.5 m from the injection site was found to be 259 µSv/h and 53.4 µSv/h, respectively, during the rest study and 301 µSv/h and 67.25 µSv/h, respectively, during stress study.
Conclusion:
The exposure to the staff performing CFR study with 13N-ammonia was well within prescribed limits by the International Commission on Radiological Protection 103. The CFR measurement with 13N-ammonia positron emission tomography/computed tomography can be included in routine workups of cardiac patients without the fear of radiation exposure.
More Related Videos
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET