Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Precordial electrocardiographic recording and QT measurement from a novel wearable ring device
G Stuart Mendenhall1, Matthew O Jones2, Charles V Pollack3
1Scripps Memorial Hospital-La Jolla, La Jolla, California.
Background:
The availability of portable and wearable electrocardiographic (ECG) devices has increased secondary to technological development. Single-lead ECG recordings have been shown to reliably detect and characterize cardiac rhythms such as atrial fibrillation. Acquisition of precordial electrodes for full 12-lead ECG reconstruction from bipolar recordings is complicated by the absence of a body ground/Wilson central terminal electrode. The extent of difference between standard precordial leads and those from a wearable bipolar ECG recorder has not been characterized.
Objective:
The purpose of this study was to characterize the precordial ECG lead set from sequential bipolar recordings from an ECG ring wearable device.
Methods:
In 70 patients who wore an ECG device on a right-hand finger, sequential precordial leads (CR1-CR6) were obtained along with chest electrodes (V1-V6). During acquisition of the modified precordial lead CR6, a full standardized 12-lead ECG capture was obtained. Signal quality was assessed using automated analysis software, and correlation values between the ring-derived ECG precordial leads and standard ECG leads were compared with regard to QRS duration, QT width, and RR interval.
Results:
High concordance in the morphologies of precordial ECG leads obtained in a standard fashion and those recorded through an ECG ring was observed. Morphologic alignment improved with increasing laterality of the precordial lead with chest to right arm ring recording (CR5, CR6) compared with anterior chest leads to right arm (CR1, CR2). Segmental measurements of QRS duration and QT segment were well aligned and of high correlation.
Conclusion:
Wearable ring-based ECG technology is capable of high-fidelity recordings of precordial leads for nonsimultaneous reconstruction of complete ECG sets. These recordings correlate highly with surface-obtained QRS and QT duration measurements and have significant implications for clinical applications. Uninterpretable tracings were primarily due to electrode noise from poor electrode contact.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...