Related Experiment Video
Updated: Aug 5, 2026

A Radio-telemetric System to Monitor Cardiovascular Function in Rats with Spinal Cord Transection and Embryonic Neural Stem Cell Grafts
Published on: October 7, 2014
Carbon Instrumentation in Patients with Metastatic Spinal Cord Compression
Søren Schmidt Morgen1, Emma Benedikte Alfthan Madsen1, Anders Skive Weiland1
1Spine Unit, Department of Orthopaedic Surgery, Rigshospitalet, University of Copenhagen, DK-2100 Copenhagen, Denmark.
Abstract:
Recently carbon spinal implants have been introduced in the treatment of patients with metastatic spinal cord compression (MSCC). This is expected to decrease the deflection of radiation and improve diagnostic imaging and radiotherapy when compared to titanium implants. The aim of this study was to determine the safety and effectiveness of spinal carbon instrumentation (CI) in patients with MSCC in a large cohort study. A total of 163 patients received instrumentation between 1 January 2017 and 31 December 2021. A total of 80 were stabilized with CI and 83 with TI. The outcome measures were surgical revision, postsurgical survival, peri-operative bleeding, and surgery time. The peri-operative blood loss in the CI-group was significantly lower than that in the TI-group: 450mL vs. 630mL, (p = 0.02). There were no significant differences between the groups in mean survival (CI 9.9) vs. (TI 12.9) months (p = 0.39), or the number of patients needing a revision (CI 6) vs. (TI 10), (p = 0.39). The median duration of surgery was 121 min, (p = 0.99) with no significant difference between the two groups. Surgical treatment with CI for MSCC is safe and an equally sufficient treatment when compared to TI.