Directed Graph Mapping exceeds Phase Mapping for the detection of simulated 2D meandering rotors in fibrotic tissue

Sebastiaan Lootens1, Iris Janssens1, Robin Van Den Abeele1

  • 1Department of Physics and Astronomy, Ghent University, Ghent, Belgium.

PubMed

Insights

Directed Graph Mapping (DGM) offers superior accuracy over Phase Mapping (PM) for detecting cardiac rotors, especially in complex conditions like high meandering, fibrosis, and noise. DGM provides a more robust solution for guiding arrhythmia ablation strategies.

Area of Science:

  • Cardiology
  • Computational Biology
  • Biomedical Engineering

Background:

  • Cardiac arrhythmias, including atrial fibrillation (AF), are often linked to re-entrant wave phenomena known as rotors.
  • Accurate detection of rotor centers is critical for effective ablation strategies in treating cardiac arrhythmias.
  • Traditional Phase Mapping (PM) methods for rotor detection are susceptible to errors, particularly in fibrotic cardiac tissue and noisy signal environments.

Purpose of the Study:

  • To compare the performance of advanced Phase Mapping (PM) techniques against a novel Directed Graph Mapping (DGM) method for detecting cardiac rotors.
  • To evaluate the robustness of both PM and DGM in simulated scenarios with varying degrees of tissue meandering, fibrosis, and temporal noise.
  • To determine which mapping technique provides more reliable rotor center identification for guiding arrhythmia ablation.

Main Methods:

  • Simulated 64 instances of 2D meandering rotors under diverse conditions of fibrosis and temporal noise.
  • Employed four comparative strategies: visual analysis, F2-score and distance distribution comparison, and mid-p McNemar statistical testing.
  • Assessed the precision and recall of rotor detection by both PM and DGM techniques across all simulations.

Main Results:

  • Both PM and DGM performed comparably well in simulations with low meandering, fibrosis, and noise.
  • In scenarios with high meandering, fibrosis, and noise, PM exhibited significant errors, primarily false positives, leading to low precision.
  • DGM demonstrated superior robustness, maintaining high F2-scores (≥0.931) compared to the best PM performance (≥0.635) across all simulations.

Conclusions:

  • Directed Graph Mapping (DGM) is a more reliable and robust technique for detecting cardiac rotors than advanced Phase Mapping (PM) methods.
  • DGM's improved performance is particularly evident in complex cardiac conditions characterized by high meandering, fibrosis, and noise.
  • DGM shows significant potential to enhance the accuracy and effectiveness of ablation strategies for treating cardiac arrhythmias like AF.