Serratia marcescens bacteremia from contaminated pressure transducers
Related Experiment Video
Updated: Aug 5, 2026

Subcutaneous Infection of Methicillin Resistant Staphylococcus Aureus (MRSA)
Published on: February 9, 2011
Abstract:
Primary Serratia marcescens bacteremia developed in 17 patients in an intensive care unit after exposure to pressure monitoring devices. A study showed that all of the transducer heads were contaminated with S marcescens, and prospective culturing of 110 pressure monitoring lines disclosed a 24% rate of contamination with the same organism. Hand contamination occurs at the time the equipment is initially assembled; in five of eight trials, transmission was shown experimentally to occur by direct inoculation of open ports. Routine disinfection of the transducer heads with glutaraldehyde not only effectively decontaminated all pressure monitoring lines in use, but also controlled the outbreak.
Insights
Serratia marcescens bacteremia outbreaks in ICUs linked to contaminated pressure monitoring devices. Routine glutaraldehyde disinfection of transducer heads effectively decontaminated lines and controlled the S marcescens spread.
Area of Science:
- Infectious Diseases
- Medical Devices
- Hospital Epidemiology
Background:
- Primary Serratia marcescens bacteremia outbreaks occurred in an intensive care unit (ICU).
- The outbreaks were associated with the use of pressure monitoring devices.
Purpose of the Study:
- To investigate the source and transmission of Serratia marcescens in ICU patients.
- To evaluate the effectiveness of glutaraldehyde disinfection in controlling the outbreak.
Main Methods:
- Culturing of transducer heads and pressure monitoring lines.
- Experimental transmission studies via direct inoculation of open ports.
- Prospective culturing of 110 pressure monitoring lines.
Main Results:
- All transducer heads tested were contaminated with S marcescens.
- A 24% contamination rate of pressure monitoring lines with S marcescens was found.
- Hand contamination during assembly and direct inoculation of open ports were identified as transmission routes.
Conclusions:
- Pressure monitoring devices are a significant source of S marcescens transmission in ICUs.
- Routine disinfection of transducer heads with glutaraldehyde is effective in decontaminating devices.
- Glutaraldehyde disinfection successfully controlled the S marcescens outbreak.