Endogenous monoclonal immunoglobulins analyzed using the EXENT® solution and LC-MS
David Barnidge1, Derek Troske1, Simon North2
1The Binding Site, Part of ThermoFisher Scientific Research and Development Laboratory, 3777 40th Ave NW, Rochester, MN 55906, United States.
Liquid chromatography-mass spectrometry (LC-MS) can detect monoclonal immunoglobulins missed by the EXENT® Solution. This reflex testing offers faster results and equivalent sensitivity for identifying these critical biomarkers.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Mass Spectrometry
Background:
- The EXENT® Solution is an automated system for identifying and quantifying monoclonal immunoglobulins in serum, combining immunoprecipitation with MALDI-TOF mass spectrometry.
- EXENT® demonstrates improved sensitivity over traditional gel-based methods like SPEP and IFE for detecting monoclonal immunoglobulins.
- The clinical significance of monoclonal immunoglobulins necessitates sensitive and accurate detection methods.
Purpose of the Study:
- To evaluate the capability of liquid chromatography-mass spectrometry (LC-MS) to detect monoclonal immunoglobulins that are not identified by the EXENT® Solution.
- To assess the feasibility of using LC-MS as a reflex test for samples negative by EXENT®.
Main Methods:
- Serum samples from patients with multiple myeloma (MM) were analyzed using the EXENT® Solution.
- Samples negative for monoclonal immunoglobulins by EXENT® were subsequently analyzed by LC-MS.
- LC-MS analysis utilized eluate from the MALDI-TOF MS spotting step, employing a low-flow LC system coupled to a Q-TOF mass spectrometer.
Main Results:
- LC-MS successfully detected monoclonal immunoglobulins in samples that tested negative by the EXENT® Solution.
- The study confirmed that LC-MS offers equivalent sensitivity to EXENT® for detecting these specific monoclonal immunoglobulins.
Conclusions:
- Reflexing samples to LC-MS for analysis is a viable strategy to identify monoclonal immunoglobulins potentially missed by EXENT®.
- This approach streamlines workflow, avoiding additional sample handling and enabling a faster time-to-result compared to other advanced methods.
- LC-MS provides a sensitive and efficient complementary method for monoclonal immunoglobulin detection in clinical diagnostics.