Related Experiment Video
Updated: Jul 2, 2025

Targeted Neuronal Injury for the Non-Invasive Disconnection of Brain Circuitry
Published on: September 27, 2020
The Safety and Efficacy of Resolving Burr Entrapment by the Ping-Pong and Mother-in-Child Techniques
Cong Thanh Nguyen1, Lam Truong Hoai2, Duc Nguyen Hung2
1Cardiovascular, Vietnam National Heart Institute, Hanoi, VNM.
Abstract:
Burr entrapment is a serious risk when performing rotational atherectomy on specific anatomical features of lesions such as tortuosity, calcification, and acute angulation. This occurrence, known as the Kokeshi phenomenon in Japanese, is caused by the burr's proximal section being unable to ablate while pulling back the burr, leaving the distal end of the burr covered in diamond crumbs capable of lesion ablation following rotation. There are reports of different approaches used to retrieve an entrapped rotablator burr. In this case, we demonstrate that the ping-pong and mother-in-child techniques, which use separate guide catheters to engage the same coronary artery wiring across the lesion afterward and deep engagement of guide extension catheter manual traction, are highly effective and secure methods for retrieval.