Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Sensitivity and specificity of handheld one lead ECG detecting atrial fibrillation in an outpatient clinic setting
Johan Malmqvist1, Johan Engdahl2, Gunnar Sjölund3
1Karolinska Institutet, Dept of Clinical Sciences, Danderyd University Hospital, Stockholm, Sweden; Dept of Clinical Physiology, Danderyd University Hospital, Stockholm, Sweden.
Insights
The Zenicor One handheld ECG demonstrated high accuracy in detecting atrial fibrillation (AF) in an outpatient setting. This device showed 98% sensitivity and 99% specificity compared to traditional 12-lead ECG.
Area of Science:
- Cardiology
- Medical Devices
Background:
- Atrial fibrillation (AF) is a common arrhythmia requiring accurate detection.
- Early diagnosis of AF is crucial for effective management and stroke prevention.
Purpose of the Study:
- To evaluate the sensitivity and specificity of the Zenicor One handheld electrocardiogram (ECG) device.
- To assess the diagnostic performance of the Zenicor One for AF detection in an outpatient clinical environment.
Main Methods:
- A prospective study involving 220 patients in outpatient clinics.
- Comparison of one-lead handheld ECG recordings with a 12-lead ECG as the gold standard.
- Dichotomous categorization of recordings as "atrial fibrillation" or "not atrial fibrillation" by senior cardiologists.
Main Results:
- The Zenicor One handheld ECG achieved 98% sensitivity in diagnosing atrial fibrillation.
- The device exhibited 99% specificity for atrial fibrillation detection.
- High agreement was observed between the handheld device and the 12-lead ECG.
Conclusions:
- The Zenicor One handheld ECG is a highly sensitive and specific tool for detecting atrial fibrillation in healthcare outpatient settings.
- This device offers a reliable alternative for AF screening and diagnosis.
- The findings support the clinical utility of the Zenicor One in routine patient care.
Aims:
To validate the sensitivity and specificity of the Zenicor One handheld ECG device for detection of atrial fibrillation in an outpatient clinical setting.
Methods And Results:
Patients attending outpatient clinics at Danderyd Hospital (n = 220) were examined with one lead handheld ECG immediately after standard care 12‑lead ECG recording. Twelve recordings were excluded (atrial flutter or pacing) or missing. The recordings were dichotomously categorized as "atrial fibrillation" or "not atrial fibrillation" by two senior cardiologists. In cases of diverging interpretations, a third senior cardiologist had the deciding vote. Sensitivity and specificity in diagnosing atrial fibrillation was calculated with 12‑lead ECG as gold standard. Sensitivity and specificity for diagnosis of atrial fibrillation with one lead handheld ECG and 12‑lead ECG as gold standard was 98% and 99% respectively.
Conclusion:
In a health-care outpatient setting, Zenicor One handheld ECG had high sensitivity and specificity for detection of atrial fibrillation when compared with 12‑lead ECG.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....