Using the Seraph® 100 Microbind® Affinity blood filter under slow flow conditions through 18 G and 16 G central lines
Malin-Theres Seffer1, Torsten M Meyer1, Dan-Nicolae Borchina1
1Medical Clinic V, Nephrology | Rheumatology | Blood Purification, Academic Teaching Hospital Braunschweig, Braunschweig, Germany.
Introduction:
The Seraph® 100 Microbind® Affinity blood filter (Seraph® 100) has been in use since 2019 for the treatment of fulminant or difficult to treat blood stream infections as an adjunct to pharmacotherapy. In 2020 the device received emergency use authorization by the US Food and Drug Administration for the treatment of critically ill COVID-19 patients with confirmed or imminent respiratory failure. Results of an international registry showed that the Seraph® 100 was operated under blood flow rates of 100-350 mL/min. As those conditions require a large bore central line, a dialysis catheter is currently considered indispensable to operate the Seraph® 100. The use of smaller catheter lumina has neither been evaluated in vitro nor in vivo.
Methods:
In vitro pressure data before and after the Seraph® 100 at various blood pump rates (prepump line 16 G, postpump line 18 G) with saline and human plasma were obtained. Further, anecdotal flow and pressure data of two patients treated with the Seraph® 100 for a COVID-19 infection are reported.
Results:
At a pump speed of 50 mL/min pre-Seraph® pressure using saline was -70 [-70 to -60] mm Hg. In comparison, using plasma pre-Seraph® pressure was lower at -120 [-120 to -105] mm Hg; p < 0.001 (t-test). The post-Seraph® pressure at 50 mL/min using saline of 120 [110-130] mm Hg was not different from plasma at 130 [120-140] mm Hg, p = 0.152 (t-test). Blood flow rates of 50 mL/min did not lead to preAP levels below -250 mm Hg in the two clinical cases.
Conclusion:
Seraph® 100 blood flow rate of 50 mL/min may be achieved using low flow vascular access, allowing to treat a blood volume 72 L in 24 h.