Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Long-term night-to-night variability of sleep-disordered breathing using a radar-based home sleep apnea test: a
Samuel Tschopp1,2, Urs Borner1, Marco Caversaccio1
1Department of Otorhinolaryngology, Head and Neck Surgery, Inselspital, University Hospital and University of Bern, Bern, Switzerland.
Study Objectives:
Night-to-night variability of sleep-disordered breathing limits the diagnostic accuracy of a single measurement. Multiple recordings using a reliable, affordable method could reduce the uncertainty and avoid misdiagnosis, which could be possible with radar-based home sleep apnea testing (HSAT).
Methods:
We recruited consecutive patients with suspected sleep-disordered breathing and performed contactless radar-based HSAT with automated scoring (Sleepiz One; Sleepiz AG, Zurich, Switzerland) over 10 nights. During the first night, patients were simultaneously measured with peripheral arterial tonometry.
Results:
Twenty-four of the 28 included patients could achieve a minimum of 4 measurements. The failure rate was 16% (37 of 238 measurements). The apnea-hypopnea index (AHI) and oxygen desaturation index were consistently lower with radar-based HSAT compared with peripheral arterial tonometry. The variability of the AHI was considerable, with a standard error of measurement of 5.2 events/h (95% confidence interval [CI]: 4.6-5.7 events/h) and a minimal detectable difference of 14.4 events/h (95% CI: 12.7-15.9 events/h). Alcohol consumption partially accounted for the variability, with an AHI increase of 1.7 events/h (95% CI: 0.6-2.8 events/h) for each standard drink. Based on a single measurement, 17% of patients were misdiagnosed and 32% were misclassified for sleep-disordered breathing severity. After 5 measurements, the mean AHI of the measured nights stabilized with no evidence of substantial changes with additional measurements.
Conclusions:
Night-to-night variability is considerable and stable over 10 nights. HSAT using radar-based methods over multiple nights is feasible and well tolerated by patients. It could offer lower costs and allow for multiple-night testing to increase accuracy. However, validation and reducing the failure rate are necessary for implementation in the clinical routine.
Clinical Trial Registration:
Registry: ClinicalTrials.gov; Name: Recording of Multiple Nights Using a New Contactless Device (Sleepiz One Connect) in Obstructive Sleep Apnea; URL: https://clinicaltrials.gov/study/NCT05134402; Identifier: NCT05134402.
Citation:
Tschopp S, Borner U, Caversaccio M, Tschopp K. Long-term night-to-night variability of sleep-disordered breathing using a radar-based home sleep apnea test: a prospective cohort study. J Clin Sleep Med. 2024;20(7):1079-1086.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Volumes and Capacities I