Performance characteristics of the 5-ring GE Discovery MI PET/CT scanner using AAPM TG-126 report
Refaat Y AlMazrou1, Shadei F Alanazi1, Mohammed H Alzaid2
1Biomedical Physics Department, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia.
Aim:
To report on the performance characteristics of the 5-ring GE Discovery MI PET/CT systems using the AAPM TG-126 report and compare these results to NEMA NU 2-2012 where applicable.
Materials And Methods:
TG-126 testing was performed on two GE 5-Rings Discovery MI scanners. Tests performed included spatial resolution, PET/CT image-registration accuracy, sensitivity, count rate performance, accuracy of corrections, image contrast, scatter/attenuation correction, and image uniformity. All acquired data were analyzed using scanner console or free software tools as described by TG-126 and the results were then compared to published NEMA NU 2-2012 values.
Results:
Both scanners gave similar resolution results for TG-126 and NEMA NU 2-2012 and were within manufacturer specifications. Image-registration accuracy between PET and CT using our clinical protocol showed excellent results with values ≤1 mm. Sensitivity using TG-126 was 19.43 cps/kBq while for NEMA the value was 20.73 cps/kBq. The peak noise-equivalent counting rate was 2174 kcps at 63.1 kBq/mL and is not comparable to NEMA NU 2-2012 due to differences in phantoms and methods used to measure and calculate this parameter. The accuracy of corrections for count losses for TG-126 were expressed in SUV values and found to be within 10% of the expected SUV measurement of 1. Image contrast and scatter/attenuation correction using the TG-126 method gave acceptable results. Image uniformity assessment resulted in values within the recommended ± 5% limits.
Conclusion:
These results show that the 5-ring GE Discovery MI PET/CT scanner testing using TG-126 is reproducible and has similar results to NEMA NU 2-2012 tests where applicable. We hope these results start to form the basis to compare PET/CT systems using TG-126.
More Related Videos
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...