Effect of reducing electrical stimulation rate on hearing performance of Nucleus® cochlear implant recipients
Komal Arora1,2, Kerrie Plant1,2, Pam Dawson1,2
1CochlearTM Limited, Melbourne, Australia.
Objective:
To evaluate whether a 500 pulses per second per channel (pps/ch) rate would provide non-inferior hearing performance compared to the 900 pps/ch rate in the Advanced Combination Encoder (ACE™) sound coding strategy.
Design:
A repeated measures single-subject design was employed, wherein each subject served as their own control. All except one subject used 900 pps/ch at enrolment. After three weeks of using the alternative rate program, both programs were loaded into the sound processor for two more weeks of take-home use. Subjective performance, preference, words in quiet, sentences in babble, music quality, and fundamental frequency (F0) discrimination were assessed using a balanced design.
Study Sample:
Data from 18 subjects were analysed, with complete datasets available for 17 subjects.
Results:
Non-inferior performance on all clinical measures was shown for the lower rate program. Subjects' preference ratings were comparable for the programs, with 53% reporting no difference overall. When a preference was expressed, the 900 pps/ch condition was preferred more often.
Conclusion:
Reducing the stimulation rate from 900 pps/ch to 500 pps/ch did not compromise the hearing outcomes evaluated in this study. A lower pulse rate in future cochlear implants could reduce power consumption, allowing for smaller batteries and processors.