Technical note: Workload and transmission data for mobile C-arm fluoroscopy in gastrointestinal endoscopy
Xinhua Li1, Theodore A Marschall1, Kai Yang1
1Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
This study analyzed mobile C-arm usage in gastrointestinal endoscopy, providing workload and radiation transmission data. Findings highlight the need for evaluating structural shielding in procedure rooms to ensure safety.
Area of Science:
- Medical Physics
- Radiological Sciences
- Endoscopic Procedures
Background:
- Mobile C-arms are utilized in fixed settings, necessitating expert evaluation of structural shielding.
- Understanding radiation exposure is crucial for patient and staff safety during fluoroscopic procedures.
Purpose of the Study:
- To assess clinical workload distributions for mobile C-arms in gastrointestinal endoscopy.
- To determine Archer equation parameters for C-arm X-ray beam spectra.
Main Methods:
- Retrospective analysis of 30 months of gastrointestinal endoscopic procedures using Cios Alpha systems.
- Examination of X-ray parameters (voltage, current-time product, air kerma, fluoroscopic time).
- Measurement of Half-Value Layer (HVL), stray radiation fraction, and Monte Carlo simulations for radiation transmission through shielding materials.
Main Results:
- 3509 procedures were analyzed, averaging 13.48 procedures per room per week.
- Dose quantities per procedure: 54.8 mGy (reference air kerma), 18.3 Gy·cm² (air kerma-area product), and 7.8 minutes (fluoroscopic time).
- Transmission fitting parameters were determined for C-arm X-ray beam voltages (90-120 kV).
Conclusions:
- The study provides essential workload and radiation transmission data for mobile C-arm fluoroscopy in GI endoscopy.
- Results indicate a clear need for structural shielding evaluation in procedure rooms to mitigate radiation exposure.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: