Technical note: Workload and transmission data for mobile C-arm fluoroscopy in gastrointestinal endoscopy
Xinhua Li1, Theodore A Marschall1, Kai Yang1
1Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
Background:
Mobile C-arms may be used in fixed locations, and it is recommended that qualified experts evaluate structural shielding.
Purpose:
To assess clinical workload distributions for mobile C-arms used in gastrointestinal endoscopy and determine the Archer equation parameters for the C-arm beam spectra.
Methods:
Consecutive (30 months) gastrointestinal endoscopic procedures on two Cios Alpha systems (Siemens) were retrospectively analyzed. X-ray tube voltage, tube current-time product, reference point air kerma (Ka,r), air kerma-area product (PKA), and fluoroscopic time were examined. The primary beam half-value layer (HVL) was measured with an ionization chamber and aluminum 1100 plates. Stray radiation fraction at 1 m from a scattering source (ACR R/F phantom) was directly measured. Monte Carlo (Geant4) simulation was performed to calculate the transmission of broad X-ray beams through lead, concrete, gypsum, and steel, with X-ray HVLs matching those of the C-arm X-ray beam. The transmission data were fitted to the Archer equation.
Results:
The number of procedures (3509) was equivalent to 13.48 procedures per room per week. Dose quantities were 54.8 mGy (Ka,r), 18.3 Gy∙cm2 (PKA), and 7.8 min (fluoroscopic time) per procedure. X-ray beam irradiation events were recorded for 2906 (82.8%) procedures with 160,009 events, whose mA-minute weighted tube voltage was 91.0 kV and the workload was 0.68 mA-minute per procedure. The two rooms had a significant difference in the number of procedures per week, 17.3 (29) [mean (maximum)] and 9.6 (16), respectively. The stray radiation fraction was 9.7×10-4 (80 kV) and 1.25×10-3 (120 kV). Transmission fitting parameters were provided for the tube voltage (on average, 90 kV; high end, 120 kV) of the C-arm.
Conclusions:
This work provides workload and transmission data for mobile C-arm fluoroscopy in gastrointestinal endoscopy, which indicates a need for structural shielding evaluation of the procedure rooms.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: