A comparison between radial artery compression devices for patent hemostasis after transradial percutaneous
Luigi Di Serafino1, Maurizio Turturo2, Claudio Larosa3
1Department of Advanced Biomedical Sciences, University of Naples Federico II, Naples, Italy.
Objectives:
Patent hemostasis (PH) is essential for preventing radial artery occlusion (RAO) after trans-radial procedures; however, it remains unclear how it should be obtained. The aim of this multicenter randomized study was to evaluate whether the use of an adjustable device (AD), inflated with a pre-determined amount of air (AoA), was more effective than a non-adjustable device (non-AD) for achieving PH, thereby reducing the incidence of RAO.
Methods:
We enrolled a total of 480 patients undergoing transradial procedure at 3 Italian institutions. Before the procedure, a modified Reverse Barbeau Test (mRBT) was performed in all patients to evaluate the AoA to be eventually inflated in the AD. After the procedure, patients were randomized into 2 groups: (1) AD Group, using TR-Band (Terumo) inflated with the pre-determined AoA; and 2) non-AD Group, using RadiStop (Abbott). An RBT was performed during compression to demonstrate the achievement of PH, as well as 24 hours later to evaluate the occurrence of RAO.
Results:
PH was more often obtained in the AD Group compared with the non-AD Group (90% vs 64%, respectively, P less than .001), with no difference in terms of bleedings. RAO occurred more often in the non-AD Group compared with the AD Group (10% vs 3%, respectively, P less than .001). Of note, mRBT was effective at guiding AD inflation and identifying high-risk patients in whom PH was more difficult to obtain.
Conclusions:
The use of AD, filled with a predetermined AoA, allowed PH significantly more often compared with non-AD, providing a significantly reduced incidence of RAO.
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...