Patient-specific non-invasive estimation of the aortic blood pressure waveform by ultrasound and tonometry
Shuran Zhou1, Kai Xu2, Yi Fang2
1College of Medicine and Biological Information Engineering, Northeastern University, Shenyang 110169, China; Aerospace Clinical Medical Center, School of Aerospace Medicine, Air Force Medical University, Xi'an 710032, China.
Insights
This study developed a new, clinically feasible method to accurately estimate aortic blood pressure (ABP) non-invasively. The technique transforms a proof-of-concept algorithm, enabling routine cardiac clinic use.
Area of Science:
- Cardiovascular Physiology
- Medical Imaging and Diagnostics
- Biomedical Engineering
Background:
- Aortic blood pressure (ABP) is a superior prognostic indicator for cardiovascular disease compared to peripheral blood pressure.
- Existing algorithms for non-invasive ABP derivation are limited to proof-of-concept due to the need for a priori ABP waveform knowledge to calculate aortic pulse wave velocity (PWV).
Purpose of the Study:
- To transform a proof-of-concept algorithm into a clinically feasible technique for non-invasive ABP estimation.
- To enable routine clinical use of accurate ABP measurements.
Main Methods:
- Utilized the Bramwell-Hill equation to non-invasively calculate aortic PWV.
- Reconstructed the ABP waveform using non-invasively determined aortic blood flow velocity, aortic diameter, and radial pressure.
- Validated the new algorithm against invasive cardiac catheterization and non-invasive SphygmoCor device measurements.
Main Results:
- Demonstrated good agreement between estimated and directly measured pressures in both invasive (e.g., SBP error: 1.4 ± 0.8 mmHg) and non-invasive (e.g., SBP error: 2.0 ± 1.4 mmHg) comparisons.
- Correlations and Bland-Altman plots further supported the accuracy of the new ABP estimation method.
Conclusions:
- The proposed method accurately estimates ABP non-invasively using standard, validated techniques.
- This technique has the potential to transition ABP estimation from research settings to routine cardiac clinic practice.
Background And Objective:
Aortic blood pressure (ABP) is a more effective prognostic indicator of cardiovascular disease than peripheral blood pressure. A highly accurate algorithm for non-invasively deriving the ABP wave, based on ultrasonic measurement of aortic flow combined with peripheral pulse wave measurements, has been proposed elsewhere. However, it has remained at the proof-of-concept stage because it requires a priori knowledge of the ABP waveform to calculate aortic pulse wave velocity (PWV). The objective of this study is to transform this proof-of-concept algorithm into a clinically feasible technique.
Methods:
We used the Bramwell-Hill equation to non-invasively calculate aortic PWV which was then used to reconstruct the ABP waveform from non-invasively determined aortic blood flow velocity, aortic diameter, and radial pressure. The two aortic variables were acquired by an ultrasound system from 90 subjects, followed by recordings of radial pressure using a SphygmoCor device. The ABPs estimated by the new algorithm were compared with reference values obtained by cardiac catheterization (invasive validation, 8 subjects aged 62.3 ± 12.7 years) and a SphygmoCor device (non-invasive validation, 82 subjects aged 45.0 ± 17.8 years).
Results:
In the invasive comparison, there was good agreement between the estimated and directly measured pressures: the mean error in systolic blood pressure (SBP) was 1.4 ± 0.8 mmHg; diastolic blood pressure (DBP), 0.9 ± 0.8 mmHg; mean blood pressure (MBP), 1.8 ± 1.2 mmHg and pulse pressure (PP), 1.4 ± 1.1 mmHg. In the non-invasive comparison, the estimated and directly measured pressures also agreed well: the errors being: SBP, 2.0 ± 1.4 mmHg; DBP, 0.8 ± 0.1 mmHg; MBP, 0.1 ± 0.1 mmHg and PP, 2.3 ± 1.6 mmHg. The significance of the differences in mean errors between calculated and reference values for SBP, DBP, MBP and PP were assessed by paired t-tests. The agreement between the reference methods and those obtained by applying the new approach was also expressed by correlation and Bland-Altman plots.
Conclusion:
The new method proposed here can accurately estimate ABP, allowing this important variable to be obtained non-invasively, using standard, well validated measurement techniques. It thus has the potential to relocate ABP estimation from a research environment to more routine use in the cardiac clinic.
Short Abstract:
A highly accurate algorithm for non-invasively deriving the ABP wave has been proposed elsewhere. However, it has remained at the proof-of-concept stage because it requires a priori knowledge of the ABP waveform to calculate aortic pulse wave velocity (PWV). This study aims to transform this proof-of-concept algorithm into a clinically feasible technique. We used the Bramwell-Hill equation to non-invasively calculate aortic PWV which was then used to reconstruct the ABP waveform. The ABPs estimated by the new algorithm were compared with reference values obtained by cardiac catheterization or a SphygmoCor device. The results showed that there was good agreement between the estimated and directly measured pressures. The new method proposed can accurately estimate ABP, allowing this important variable to be obtained non-invasively, using standard, well validated measurement techniques. It thus has the potential to relocate ABP estimation from a research environment to more routine use in the cardiac clinic.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.