Related Experiment Video
Updated: Jun 26, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Magnetically driven antegrade intramedullary lengthening nails for tibial lengthening
Bjoern Vogt1, Milena Lueckingsmeier1, Georg Gosheger2
1Paediatric Orthopaedics, Deformity Reconstruction and Foot Surgery, Muenster University Hospital, Muenster, Germany.
Aims:
As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach.
Methods:
A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30).
Results:
A median tibial distraction of 44 mm (IQR 31 to 49) was achieved with a mean distraction index of 0.5 mm/day (standard deviation 0.13) and median consolidation index of 41.2 days/cm (IQR 34 to 51). Accuracy, precision, and reliability were 91%, 92%, and 97%, respectively. New temporary range of motion limitations occurred in 51% of segments (34/67). Distraction-related equinus deformity treated by Achilles tendon lengthening was the most common major complication recorded in 16% of segments (11/67). In 95% of patients (55/58) the distraction goal was achieved with 42% unplanned additional interventions per segment (28/67). The median postoperative LD-SRS-30 score was 4.0 (IQR 3.6 to 4.3).
Conclusion:
Tibial distraction osteogenesis using motorized ILNs inserted via an antegrade approach appears to be a reliable and precise procedure. Temporary joint stiffness of the knee or ankle should be expected in up to every second patient. A high rate and wide range of complications of variable severity should be anticipated.