A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
26.2K
Safety and performance of a novel subxiphoidal pacemaker system.
Imad Libbus1, Venkatakrishna Tholakanahalli2, Henri Roukoz2
1Calyan Technologies, Saint Paul, MN, USA. imad@calyantech.com.
Summary
This study shows a new leadless pacemaker implanted outside the heart is safe and works for short-term pacing in swine. Further research is needed for human use.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Traditional pacemakers carry risks due to intracardiac leads.
- Leadless and extravascular pacemakers offer potential to reduce complications and improve access.
- Novel pacing technologies are crucial for advancing cardiac care.
Purpose of the Study:
- To evaluate the implantation, short-term safety, and performance of a new subxiphoidal extracardiac pacemaker.
- To assess the feasibility of a leadless pacing solution placed beneath the sternum.
- To investigate the potential of extravascular pacing for cardiac rhythm management.
Main Methods:
- 16 Yorkshire Cross swine underwent implantation of the subxiphoidal pacemaker.
- The device was placed under the sternum with the electrode on the anterior pericardium.
- Animals were monitored for 90 days, with regular measurements of pacing capture threshold and impedance.
Main Results:
- Successful implantation was achieved in all 16 animals.
- Consistent initial pacing capture threshold (PCT) and impedance were recorded.
- PCT showed a slight increase and stabilized, with stable impedance over 90 days.
- No adverse events affecting animal welfare were observed.
Conclusions:
- The novel subxiphoidal extracardiac pacemaker demonstrated safety and feasibility for short-term extravascular pacing.
- This leadless device shows promise as an alternative to traditional pacemakers.
- Further human studies are necessary to confirm long-term safety and efficacy.