Technical assessment of resolution of handheld ultrasound devices and clinical implications
Moritz Herzog1, Maia Arsova2, Katja Matthes2
1Else Kröner Fresenius Center for Digital Health, TU Dresden Faculty of Medicine Carl Gustav Carus, Dresden, Germany.
Purpose:
Since handheld ultrasound devices are becoming increasingly ubiquitous, objective criteria to determine image quality are needed. We therefore conducted a comparison of objective quality measures and clinical performance.
Material And Methods:
A comparison of handheld devices (Butterfly IQ+, Clarius HD, Clarius HD3, Philips Lumify, GE VScan Air) and workstations (GE Logiq E10, Toshiba Aplio 500) was performed using a phantom. As a comparison, clinical investigations were performed by two experienced ultrasonographers by measuring the resolution of anatomical structures in the liver, pancreas, and intestine in ten subjects.
Results:
Axial full width at half maximum resolution (FWHM) of 100µm phantom pins at depths between one and twelve cm ranged from 0.6-1.9mm without correlation to pin depth. Lateral FWHM resolution ranged from 1.3-8.7mm and was positively correlated with depth (r=0.6). Axial and lateral resolution differed between devices (p<0.001) with the lowest median lateral resolution observed in the E10 (5.4mm) and the lowest axial resolution (1.6mm) for the IQ+ device. Although devices showed no significant differences in most clinical applications, ultrasonographers were able to differentiate a median of two additional layers in the wall of the sigmoid colon and one additional structure in segmental portal fields (p<0.05) using cartwheel devices.
Conclusion:
While handheld devices showed superior or similar performance in the phantom and routine measurements, workstations still provided superior clinical imaging and resolution of anatomical substructures, indicating a lack of objective measurements to evaluate clinical ultrasound devices.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):