Effect of pacing site on P wave parameters: Within-patient comparison of right atrial appendage and Bachmann's bundle
Natasha A Vedage1, Edmond M Cronin1
1Section of Cardiology, Department of Medicine, Lewis Katz School of Medicine at Temple University, Philadelphia, PA, United States of America.
Background:
Conventional right atrial appendage pacing (RAAp) is associated with adverse clinical outcomes mediated in part by electromechanical atrial delays. Bachmann's bundle pacing (BBp) offers more physiologic atrial activation; however, detailed analysis of pacing site on paced P wave parameters is lacking.
Methods:
Intraprocedural electrocardiograms of 21 consecutive patients undergoing atrial lead implantation were retrospectively analyzed and within-patient comparisons of 7 P wave parameters (P wave duration, P wave voltage, P wave area, PR interval, PR segment, PTFV1 and P wave axis) during sinus rhythm, RAAp and BBp performed.
Results:
The median basal P wave duration was prolonged at 134.5 ms (Q1,Q3: 120.5, 150.5) and similarly prolonged during RAAp at 144.0 ms (127.0, 176.0) but was significantly reduced with BBp at 98.0 ms (93.0, 116.0; p = 0.005 and p < 0.001, respectively). The median basal P wave voltage in lead II was normal at 0.11 mV (0.08, 0.15) but significantly reduced during RAAp at 0.08 mV (0.04, 0.11) and greatest during BBp at 0.16 mV (0.09, 0.19; p < 0.001 and p = 0.003, respectively). The median basal PR interval was top normal at 185.0 ms (163.0, 213.0) and similarly prolonged during RAAp at 204.0 ms (166.5, 221.0) but was significantly shortened during BBp at 163.0 ms (142.0, 208.0; p = 0.03 and p = 0.001, respectively).
Conclusions:
BBp has favorable effects on the paced P wave parameters including marked shortening in P wave duration, increase in P wave voltage in lead II and increase in PR segment which may offer significant hemodynamic advantages over conventional RAAp.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:39Rat Model of Right-Sided Cardiac Remodeling and Arrhythmia Using Pulmonary Artery Banding
Published on: August 30, 2024
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Electrophysiology of Normal Cardiac Rhythm
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation