A Novel, Wearable Inertial Measurement Unit for Stroke Survivors: Validity, Acceptability, and Usability
Lauri Bishop1, Marika Demers2, Justin Rowe3
1Division of Biokinesiology and Physical Therapy, Herman Ostrow School of Dentistry, University of Southern California, Los Angeles, CA; Department of Physical Therapy, Miller School of Medicine, University of Miami, Coral Gables, FL.
Objective:
To establish the concurrent validity, acceptability, and sensor optimization of a consumer-grade, wearable, multi-sensor system to capture quantity and quality metrics of mobility and upper limb movements in stroke survivors.
Design:
Single-session, cross-sectional.
Setting:
Clinical research laboratory.
Participants:
Thirty chronic stroke survivors (age 57 (10) years; 33% female) with mild to severe motor impairments participated.
Interventions:
Not Applicable.
Main Outcome Measures:
Participants donned 5 sensors and performed standardized assessments of mobility and upper limb (UL) movement. True/false, positive/negative time in active movement for the UL were calculated and compared to criterion-standards using an accuracy rate. Bland-Altman plots and linear regression models were used to establish concurrent validity of UL movement counts, step counts, and stance time symmetry of MiGo against established criterion-standard measures. Acceptability and sensor optimization were assessed through an end-user survey and decision matrix.
Results:
Mobility metrics showed excellent association with criterion-standards for step counts (video: r=0.988, P<.001, IMU: r=0.921, P<.001) and stance-time symmetry (r=0.722, P<.001). In the UL, movement counts showed excellent to good agreement (paretic: r=0.849, P<.001, nonparetic: r=0.672, P<.001). Accuracy of active movement time was 85.2% (paretic) and 88.0% (nonparetic) UL. Most participants (63.3%) had difficulty donning/doffing the sensors. Acceptability was high (4.2/5).
Conclusions:
The sensors demonstrated excellent concurrent validity for mobility metrics and UL movements of stroke survivors. Acceptability of the system was high, but alternative wristbands should be considered.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...