Non-invasive Assessment of Microvascular and Endothelial Function
Published on: January 29, 2013
MEMS Fingertip Strain Plethysmography for Cuffless Estimation of Blood Pressure
Objective:
To develop a cuffless method for estimating blood pressure (BP) from fingertip strain plethysmography (SPG) recordings.
Methods:
A custom-built micro-electromechanical systems (MEMS) strain sensor is employed to record heartbeat-induced vibrations at the fingertip. An XGboost regressor is then trained to relate SPG recordings to beat-to-beat systolic BP (SBP), diastolic BP (DBP), mean arterial pressure (MAP) values. For this purpose, each SPG segment in this setup is represented by a feature vector consisting of cardiac time interval, amplitude features, statistical properties, and demographic information of the subjects. In addition, a novel concept, coined geometric features, are introduced and incorporated into the feature space to further encode the dynamics in SPG recordings. The performance of the regressor is assessed on 32 healthy subjects through 5-fold cross-validation (5-CV) and leave-subject-out cross validation (LSOCV).
Results:
Mean absolute errors (MAEs) of 3.88 mmHg and 5.45 mmHg were achieved for DBP and SBP estimations, respectively, in the 5-CV setting. LSOCV yielded MAEs of 8.16 mmHg for DBP and 16.81 mmHg for SBP. Through feature importance analysis, 3 geometric and 26 integral-related features introduced in this work were identified as primary contributors to BP estimation. The method exhibited robustness against variations in blood pressure level (normal to critical) and body mass index (underweight to obese), with MAE ranges of [1.28, 4.28] mmHg and [2.64, 7.52] mmHg, respectively.
Conclusion:
The findings suggest high potential for SPG-based BP estimation at the fingertip.
Significance:
This study presents a fundamental step towards the augmentation of optical sensors that are susceptible to dark skin tones.
More Related Videos
09:56Implantation of Combined Telemetric ECG and Blood Pressure Transmitters to Determine Spontaneous Baroreflex Sensitivity in Conscious Mice
Published on: February 14, 2021
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Assessing Blood pressure in the Leg
Preparation: