Melanin bias in pulse oximetry explained by light source spectral bandwidth
Andrew Bierman1, Kevin Benner2, Mark S Rea2
1Namreiba LLC, Albany, NY, USA.
Background:
Pulse oximetry uses noninvasive optical measurements of light transmission from each of two sources through vascularised living tissue over the cardiac cycle (SpO2). From those measurements, the relative amount of oxygenated haemoglobin (SaO2) in circulating blood can be deduced. Recent reports have shown that, compared with SaO2 measurements from blood samples, SpO2 measurements are biased erroneously high for patients with dark skin.
Methods:
We developed a new method, spectrally resolved photoplethysmography (srPPG), to examine how spectral bandwidth affects the transmission of polychromatic light through the fingertip across the cardiac cycle. We measured and recorded the spectral transmission through the fingertip as the O2 concentration in inspired air was reduced. We applied digital spectral filters of two different bandwidths, narrow or broad, to the same srPPG recordings to determine whether SpO2 readings systematically varied for the two bandwidths. The srPPG method also allowed us to measure the fractional amount of melanin in the optical path. The effect of melanin content on the ratio of SpO2 readings for narrow and broad spectral bandwidths was analysed.
Results:
We hypothesised, based upon the Beer-Lambert law, and then showed experimentally, that the light emission spectra of light-emitting diode light sources, as used in commercial pulse oximeters, result in erroneously high SpO2 measurements for patients having greater melanin concentrations in their skin than those of the subject pool used for instrument calibration.
Conclusions:
To eliminate melanin bias, pulse oximeters should use much narrower spectral bandwidths than those used in current models.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pigmentation
Melanin occurs in two primary forms: eumelanin that provides black and brown pigment and pheomelanin that provides red color. Dark-skinned individuals produce more melanin than those with pale...
Photoreceptors and Visual Pathways
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Changes in Skin Color: Clinical Perspectives
Albinism
Albinism is a genetic disorder that affects (completely or partially) the coloring of skin, hair, and eyes. The defect is primarily...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...