Related Experiment Video
Updated: Jul 1, 2025

The Intra-Aortic Balloon Pump
Published on: February 5, 2021
A complication of an axillary intra-aortic balloon pump
Eileen Kodack1, Riya Patel2, Carain Bonner1
1Department of Anesthesiology and Perioperative Medicine, Robert Wood Johnson Medical School, New Brunswick, NJ 08901, United States.
Abstract:
Intra-aortic balloon pumps (IABPs) are used to mechanically temporize a failing heart by decreasing afterload while increasing coronary perfusion pressure of the heart while more definitive treatment is sought. We report a case of a 65-year-old male with nonischemic cardiomyopathy, atrial fibrillation, thyroiditis, and non-Hodgkin lymphoma who presented with worsening heart failure. He underwent a percutaneous placement of a left axillary IABP with seemingly no complications. Approximately 3 weeks post-placement, the patient was taken for a heart transplant when an intraoperative transesophageal echo showed that the IABP was in the aortic arch and ascending aorta, instead of its proper placement in the descending aorta. The patient's arterial line showed waveforms appropriate for an IABP patient, and the patient showed no signs indicative of improper placement. This erroneous placement carried the potential to affect the aortic valve function, injure the aortic intima and/or occlude the aortic arch vessels. .