Subcutaneous air entrapment after subcutaneous implantable cardioverter defibrillator implantation evaluated by
Yuka Taguchi1, Toshiyuki Ishikawa1, Katsumi Matsumoto1
1Department of Cardiology, Yokohama City University Graduate School of Medicine, Yokohama-City, Japan.
Background:
Inappropriate shock (IAS) caused by subcutaneous air entrapment (AE) in an early period after subcutaneous implantable cardioverter defibrillator (S-ICD) implantation has been reported, however, no detailed data on air volume are available. We evaluated the subcutaneous air volume after implantation and its absorption rate one week after implantation.
Methods:
Patients who underwent S-ICD implantation in our hospital received chest CT scans immediately after implantation and followed up 1 week later. The total subcutaneous air volume, air around the generator, the distal electrode, and the proximal electrode within 3 cm were calculated using a three-dimensional workstation. Fat areas at the level of the lower edge of the generator were also analyzed.
Result:
Fifteen patients received CT immediately after implantation. The mean age was 45.6 ± 17.9 (66.7% of men), and the mean body mass index was 24.3 ± 3.3. The three-incision technique was applied in seven patients and two-incision technique was in the latter eight patients. The mean total subcutaneous air volume was 18.54 ± 7.50 mL. Air volume around the generator, the distal electrode, and the proximal electrode were 11.05 ± 5.12, 0.72 ± 0.72, and 0.88 ± 0.87 mL, respectively. Twelve patients received a follow-up CT 1 week later. The mean total subcutaneous air was 0.25 ± 0.45 mL, showing a 98.7% absorption rate.
Conclusion:
Although subcutaneous air was observed in all patients after S-ICD implantation, most of the air was absorbed within 1 week, suggesting a low occurrence of AE-related IAS after a week postoperation.