Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Comparison of tracheal versus esophageal temperatures during laparoscopic surgery
Eun Kyung Lee1, Yu Jeong Bang2, Jeayoun Kim2
1Department of Anesthesiology and Pain Medicine, Chung-Ang University College of Medicine, Seoul, Republic of Korea.
Purpose:
Recently, endotracheal tubes with an embedded temperature sensor in the inner surface of the tube cuff (temperature tracheal tubes) have been developed. We sought to assess whether temperature tracheal tubes show a good agreement with esophageal temperature probes during surgery.
Methods:
We enrolled 40 patients who underwent laparoscopic surgery in an observational study. The tracheas of all patients were intubated with a temperature tracheal tube, and an esophageal temperature probe was inserted into the esophagus. Tracheal and esophageal temperatures were recorded at 15-min intervals until the end of surgery. Temperatures from both devices were analyzed using Bland-Altman analysis, four-quadrant plots, and polar plots.
Results:
We analyzed 261 data points from 36 patients. Temperatures ranges were 34.2 °C to 36.6 °C for the tracheal temperature tube and 34.7 °C to 37.2 °C for the esophageal temperature probe. Bland-Altman analysis showed an acceptable agreement between the two devices, with an overall mean bias (95% limit of agreement) of -0.3 °C (-0.8 °C to 0.1 °C) and a percentage error of 3%; the trending ability (temperature changes over time) between the two devices showed a concordance rate of 94% in four-quadrant plot (cut-off ≥ 92%), but this was higher than the acceptable mean angular bias of 177° (cut-off < ± 5°) and radial limits of agreement of 52° (cut-off < ± 30°) in the polar plot. Bronchoscopy during extubation and patient interviews at six hours postoperatively revealed no serious injuries related to the use of the temperature tracheal tube.
Conclusion:
The temperature tracheal tube showed an acceptable overall mean bias of -0.3 °C and a percentage error of 3%, but incompatible trending ability with the esophageal temperature probe.
Study Registration:
cris.nih.go.kr (KCT0007265); 22 April 2022.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
09:04Low-Cost Single-Port LoCoSP Device for a Transcervical Approach in Minimally Invasive Transhiatal Esophagectomy
Published on: September 11, 2021
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...