No gadolinium K-edge detected on the first clinical photon-counting computed tomography scanner
Erik Baubeta1,2, Eva Laurin Gadsböll1, Leon Will1
1Department of Imaging and Functional Medicine, Skåne University Hospital, Lund, Sweden.
Purpose:
This study aimed to elucidate whether gadolinium contrast in clinically relevant doses can be used with photon-counting computed tomography (PCCT) as an alternative contrast agent in clinical applications.
Material/Methods:
A CTDI phantom with 3D printed rods filled with different concentrations of gadolinium and iodine contrast was scanned in a PCCT and an energy-integrated computed tomography (EICT). Attenuation values at different monoenergetic steps were extracted for each contrast concentration.
Results:
For PCCT, gadolinium reached an attenuation >100 HU (103 HU) at 40 keV with a concentration 5 mmol/L whereas the same level was reached at 50 keV (118 HU) for 10 mmol/L and 90 keV (114 HU) for 25 mmol/L. For iodine, the same level of attenuation was reached at 100 keV (106 HU) with a concentration 8.75 mg I/mL. For EICT the lowest gadolinium contrast concentration needed to reach >100 HU (108 HU) was 10 mmol/L at 50 keV. For 25 mmol/L 100 HU was reached at 100 keV. For iodine contrast 108 HU was reached at 110 keV for 8.75 mg I/mL.
Conclusion:
No K-edge potential or difference in attenuation curves between iodine and gadolinium contrast is detected on the first clinical available PCCT. Clinically relevant attenuation levels were barely achieved in this setting with gadolinium concentrations approved for human use. The results of this study suggest that, given current scanning technology, gadolinium is not a clinically useful contrast agent for computed tomography because no K-edge was detected.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Magnetic Resonance Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET