Related Experiment Video
Updated: Jul 1, 2025

10:10
Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
8.0K
Linker-Preserved Iron Metal-Organic Framework-Based Lateral Flow Assay for Sensitive Transglutaminase 2 Detection in
Mulya Supianto1, Dong Kyu Yoo1, Hagyeong Hwang2
1Department of Chemistry and Green-Nano Materials Research Center, Kyungpook National University, 80 Daehakro, Buk-gu, Daegu-city 41566, Republic of Korea.
ACS Sensors
|March 12, 2024
Summary
This study introduces a novel metal-organic framework, MIL-101(Fe), as a highly sensitive label for detecting transglutaminase 2 (TGM2), a biomarker for chronic kidney disease (CKD). This advancement offers a rapid, naked-eye detection method for TGM2 in urine samples.
Area of Science:
- Materials Science
- Biomedical Engineering
- Analytical Chemistry
Background:
- Chronic kidney disease (CKD) diagnosis relies on identifying specific biomarkers.
- Transglutaminase 2 (TGM2) is a recognized biomarker for CKD.
- Existing detection methods for TGM2 can be limited in sensitivity and speed.
Purpose of the Study:
- To develop a highly sensitive and rapid colorimetric lateral flow assay (LFA) for TGM2 detection.
- To utilize the metal-organic framework MIL-101(Fe) as a novel labeling material for LFAs.
- To establish a visual, naked-eye detection method for TGM2 in urine specimens.
Main Methods:
- Surface modification of MIL-101(Fe) with oxalyl chloride, adipoyl chloride, and poly(acrylic) acid (PAA).
- Development of a colorimetric lateral flow assay (LFA) incorporating the MIL-101(Fe)-PAA label.
- Validation of the LFA's sensitivity, reproducibility, recovery, and storage stability.
- Application of machine learning for precise TGM2 concentration determination.
Main Results:
- The MIL-101(Fe)-PAA labeled LFA achieved detection limits as low as 0.009 nM for TGM2 in urine.
- This represents a 55-fold improvement in sensitivity compared to conventional gold nanoparticle (AuNP)-based LFAs.
- The assay demonstrated excellent reproducibility (CV < 2.9%), recovery (95.9-103.2%), and storage stability (CV < 6.0% over 25 days).
- Detection of TGM2 was achieved within 20 minutes.
Conclusions:
- MIL-101(Fe)-PAA serves as an effective and sensitive visual label for colorimetric LFAs.
- The developed LFA provides a rapid, sensitive, and reliable method for TGM2 detection in urine.
- This assay holds significant potential for the clinical diagnosis and monitoring of CKD.
Keywords:
lateral flow assaymachine learningmetal−organic frameworknaked-eye detectiontransglutaminase 2urine sample