Eddy Current Position Measurement in Harsh Environments: A Temperature Compensation and Calibration Approach
Gabriel Gruber1, Bernhard Schweighofer1, Matthias Berger2
1Christian Doppler Laboratory for Measurement Systems for Harsh Operating Conditions, Institute of Electrical Measurement and Sensor Systems, Graz University of Technology, 8010 Graz, Austria.
Abstract:
Eddy current displacement sensors (ECDSs) are widely used for the noncontact position measurement of small displacements (lift-offs). Challenges arise with larger displacements as the sensitivity of the ECDSs decreases. This leads to a more pronounced impact of temperature variations on the inductance and, consequently, an increased position error. Design solutions often rely on multiple coils, suitable coil carrier materials, and compensation measures to address the challenges. This study presents a single-coil ECDS for large displacement ranges in environments with high temperatures and temperature variations. The analysis is based on a sensor model derived from an equivalent circuit model (ECM). We propose design measures for both the sensing coil and the target, focusing on material selection to handle the impact of temperature variations. A key part of improving performance under varying temperatures includes model-based temperature compensation for the inductance of the sensing coil. We introduce a method to calibrate the sensor for large displacements, using a modified coupling coefficient based on field simulation data. Our analysis shows that this single-coil ECDS design maintains a position error of less than 0.2% full-scale for a temperature variation of 100 K for the sensing coil and 110 K for the target.
More Related Videos
06:17Quantifying the Relative Thickness of Conductive Ferromagnetic Materials Using Detector Coil-Based Pulsed Eddy Current Sensors
Published on: January 16, 2020
06:33Method for Simultaneous fMRI/EEG Data Collection during a Focused Attention Suggestion for Differential Thermal Sensation
Published on: January 5, 2014
Related Concept Videos
Electronic Distance Measuring Instruments
Distance Corrections
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Common Leveling Mistakes and Errors
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...