Related Experiment Video
Updated: May 6, 2026

Non-invasive Optical Imaging of the Lymphatic Vasculature of a Mouse
Published on: March 8, 2013
Near-Infrared Fluorescence Imaging Sensor with Laser Diffuser for Visualizing Photoimmunotherapy Effects under
Toshihiro Takamatsu1,2, Hideki Tanaka3, Tomonori Yano1,4
1Exploratory Oncology Research & Clinical Trial Center, National Cancer Center, Kashiwa 277-8577, Chiba, Japan.
Abstract:
The drug efficacy evaluation of tumor-selective photosensitive substances was expected to be enabled by imaging the fluorescence intensity in the tumor area. However, fluorescence observation is difficult during treatments that are performed during gastrointestinal endoscopy because of the challenges associated with including the fluorescence filter in the camera part. To address this issue, this study developed a device that integrates a narrow camera and a laser diffuser to enable fluorescence imaging through a forceps port. This device was employed to demonstrate that a laser diffuser with an NIR fluorescence imaging sensor could be delivered through a 3.2 mm diameter port. In addition, fluorescence images of Cetuximab-IR700 were successfully observed in two mice, and the fluorescence intensity confirmed that the fluorescence decayed within 330 s. This device is expected to have practical application as a tool to identify the optimal irradiation dose for tumor-selective photosensitive substances under endoscopy.