Qualitative and quantitative image quality of coronary CT angiography using photon-counting computed tomography:
Borbála Vattay1, Melinda Boussoussou1, Milán Vecsey-Nagy1
1Heart and Vascular Center, Semmelweis University, 1122 Budapest, Városmajor Street 68., Hungary.
Purpose:
We aimed to identify the optimal reconstruction settings based on qualitative and quantitative image quality parameters on standard and ultra-high resolution (UHR) images using photon-counting CT (PCCT).
Method:
We analysed 45 patients, 29 with standard and 16 with UHR acquisition, applying both smoother and sharper kernel settings. Coronary CT angiography images were performed on a dual-source PCCT system using standard (0.4/0.6 mm slice thickness, Bv40/Bv44 kernels, QIR levels 0-4) or UHR acquisition (0.2/0.4 mm slice thickness, Bv44/Bv56 kernels, QIR levels 0-4). Qualitative image quality was assessed using a 4-point Likert scale. Image noise (SD), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated in both the proximal and distal segments.
Results:
On standard resolution, larger slice thickness resulted in an average increase of 12.5 % in CNR, whereas sharper kernel led to an average 8.7 % decrease in CNR. Highest CNR was measured on 0.6 mm, Bv40, QIR4 images and lowest on 0.4 mm, Bv44, QIR0 images: 25.8 ± 4.1vs.8.3 ± 1.6 (p < 0.001). On UHR images, highest CNR was observed on 0.4 mm, Bv40, QIR4 and lowest on 0.2 mm, Bv56 and QIR0 images: 21.5 ± 3.9vs.3.6 ± 0.8 (p < 0.001). Highest qualitative image quality was found on images with Bv44 kernel and QIR level 3/4 with both slice thicknesses on standard reconstruction. Additionally, Bv56 with QIR4 on 0.2 mm slice thickness images showed highest subjective image quality. Preserved distal vessel visualization was detected using QIR 2-4, Bv56 and 0.2 mm slice thickness.
Conclusions:
Photon-counting CT demonstrated high qualitative and quantitative image quality for the assessment of coronaries and stents.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...