Six SIGMA evaluation of 17 biochemistry parameters using bias calculated from internal quality control and external
Tülay Çevlik1, Goncagül Haklar1
1Marmara University Pendik E&R Hospital, Biochemistry Laboratory, Istanbul, Turkey.
Background:
Six Sigma is a popular quality management system that enables continuous monitoring and improvement of analytical performance in the clinical laboratory. We aimed to calculate sigma metrics and quality goal index (QGI) for 17 biochemical analytes and compare the use of bias from internal quality control (IQC) and external quality assurance (EQA) data in the calculation of sigma metrics.
Methods:
This retrospective study was conducted in Marmara University Pendik E&R Hospital Biochemistry Laboratory. Sigma metrics calculation was performed as (TEa-bias)/CV). CV was calculated from IQC data from June 2018 - February 2019. EQA bias was calculated as the mean of % deviation from the peer group means in the last seven surveys, and IQC bias was calculated as (laboratory control result mean-manufacturer control mean)/ manufacturer control mean) x100. In parameters where sigma metrics were <5; QGI=bias/1.5 CV) score of <0.8 indicated imprecision, >1.2 pointed inaccuracy, and 0.8-1.2 showed both imprecision and inaccuracy.
Results:
Creatine kinase (both levels), iron and magnesium (pathologic levels) showed an ideal performance with ≥6 sigma level for both bias determinations. Eight of the 17 parameters had different sigma levels when we compared sigma values calculated from EQA and IQC derived bias% while the rest were grouped at the same levels.
Conclusions:
Sigma metrics is a good quality tool to assess a laboratory's analytical performance and facilitate the comparison of the assay performances in the same manner across multiple systems. However, we might need to design a tight internal quality control protocol for analytes showing poor assay performance.
More Related Videos
19:16The Importance of Correct Protein Concentration for Kinetics and Affinity Determination in Structure-function Analysis
Published on: March 17, 2010
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
Data Validation
Key parameters for method validation include:
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...