Fe-MOF/AuNP-based ratiometric electrochemical immunosensor for the detection of deoxynivalenol in grain products
Zaixi Shu1,2, Huilin Hu1,2, Zhenhong Yuan1
1College of Food Science and Engineering, Wuhan Polytechnic University, Wuhan, 430023, China.
Abstract:
A ratiometric assay was designed to improve the sensitivity and reliability of electrochemical immunosensors for deoxynivalenol (DON) detection. The indicator signal caused by the Fe-based metal-organic framework nanocomposites loaded with gold nanoparticles and the internal reference signal from the [Fe(CN)6]3-/4- in the electrolyte came together at the immunosensor. When immunoreactivity occurred, the indicator signals decreased as the concentration of DON increased, while the internal reference signals increased slightly. The ratio of the indicator signal to the internal reference signal was available for reproducible and sensitive monitoring of DON. The prepared immunosensor showed excellent performance in the range from 0.5 to 5000 pg mL-1, and the detection limit was 0.0166 pg mL-1. The immunosensor achieved satisfactory detection toward DON in spiked and actual samples and has a promising application in the control of DON in grain products.