Evaluation of pulse oximeter at the nasal septum during general anesthesia: comparison with finger oximeter
Yena Oh1, Duk Kyung Kim2, Dae Kyun Ryu3
1Dpartment of Anesthesiology and Pain Medicine, CHA Bundang Medical Center, CHA University, Seongnam, Republic of Korea.
Purpose:
Though the finger is generally recommended for pulse oxygen saturation (SpO2) monitoring site, its reliability may be compromised in conditions of poor peripheral perfusion. Therefore, we compared the performance of nasal septum SpO2 monitoring with finger SpO2 monitoring relative to simultaneous arterial oxygen saturation (SaO2) monitoring in generally anesthetized patients.
Methods:
In 23 adult patients, comparisons of SpO2 measured at the nasal septum and finger with simultaneous SaO2 were made at four time points during the 90 min study period. A pulse oximetry monitoring failure was defined as a > 10 s continuous failure of in an adequate SpO2 data acquisition. Core temperature as well as finger-tip and nasal septum temperatures were simultaneously measured at 10 min intervals.
Results:
A total of 92 sets of SpO2 and SaO2 measurements were obtained in 23 patients. The bias and precision for SpO2 measured at the nasal septum were - 0.8 ± 1.3 (95% confidence interval: - 1.1 to - 0.6), which was similar to those for SpO2 measured at the finger (- 0.6 ± 1.4; 95% confidence interval: - 0.9 to - 0.4) (p = 0.154). Finger-tip temperatures were consistently lower than other two temperatures at all time points (p < 0.05), reaching 33.5 ± 2.3 °C at 90 min after induction of anesthesia. While pulse oximetry monitoring failure did not occur for nasal septum probe, two cases of failure occurred for finger probe.
Conclusions:
Considering the higher stability to hypothermia with a similar accuracy, nasal septum pulse oximetry may be an attractive alternative to finger pulse oximetry. Trail registration This study was registered with Clinical Research Information Service (CRIS: https://cris.nih.go.kr/cris/en/ ; ref: KCT0008352).
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: