Resolving Water, Proteins, and Lipids from In Vivo Confocal Raman Spectra of Stratum Corneum through a Chemometric Approach
Published on: September 26, 2019
Biochemical component analysis of human myopic corneal stroma using the Raman spectrum
Jing Li1, Zhe Yu1, Ming Han2,3
1Shaanxi Eye Hospital, Xi'an People's Hospital (Xi'an Fourth Hospital, Affliated People's Hospital of Northwest University, No 4. Jiefang Road, Xin-Chen District, Xi'an, 710004, Shaanxi, China.
Purpose:
This study aimed to measure the Raman spectrum of the human corneal stroma lens obtained from small incision lenticule extraction surgery (SMILE) in Asian myopic eyes using a confocal Raman micro-spectrometer built in the laboratory.
Methods:
Forty-three myopic patients who underwent SMILE with equivalent diopters between - 4.00 and - 6.00 D were selected, and the right eye data were collected. Corneal stroma lenses were obtained during surgery, and the Raman spectra were measured after air drying. The complete Raman spectrum of human myopic corneal stroma lens tissue was obtained within the range of 700-4000 cm-1.
Results:
Thirteen characteristic peaks were found, with the stronger peaks appearing at 937 cm-1, corresponding to proline, valine, and the protein skeleton of the human myopic corneal stroma lens; 1243 cm-1, corresponding to collagen protein; 1448 cm-1, corresponding to the collagen protein and phospholipids; and 2940 cm-1, corresponding to the amino acid and lipids, which was the strongest Raman peak.
Conclusion:
These results demonstrated that Raman spectroscopy has much potential as a fast, cost-effective, and reliable diagnostic tool in the diagnosis and treatment of eye diseases, including myopia, keratoconus, and corneal infection.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...