A Computational Investigation of the Effects of Temporal Synchronization of Left Ventricular Assist Device Speed
Angela Straccia1, Fanette Chassagne2, Michael C Barbour3
1Department of Mechanical Engineering, University of Washington, Seattle, WA, USA. straccia@uw.edu.
Abstract:
Patients with advanced heart failure are implanted with a left ventricular assist device (LVAD) as a bridge-to-transplantation or destination therapy. Despite advances in pump design, the risk of stroke remains high. LVAD implantation significantly alters intraventricular hemodynamics, where regions of stagnation or elevated shear stresses promote thrombus formation. Third generation pumps incorporate a pulsatility mode that modulates rotational speed of the pump to enhance in-pump washout. We investigated how the timing of the pulsatility mode with the cardiac cycle affects intraventricular hemodynamic factors linked to thrombus formation. Computational fluid dynamics simulations with Lagrangian particle tracking to model platelet behavior in a patient-specific left ventricle captured altered intraventricular hemodynamics due to LVAD implantation. HeartMate 3 incorporates a pulsatility mode that modulates the speed of the pump every two seconds. Four different timings of this pulsatility mode with respect to the cardiac cycle were investigated. A strong jet formed between the mitral valve and inflow cannula. Blood stagnated in the left ventricular outflow tract beneath a closed aortic valve, in the near-wall regions off-axis of the jet, and in a large counterrotating vortex near the anterior wall. Computational results showed good agreement with particle image velocimetry results. Synchronization of the pulsatility mode with peak systole decreased stasis, reflected in the intraventricular washout of virtual contrast and Lagrangian particles over time. Temporal synchronization of HeartMate 3 pulsatility with the cardiac cycle reduces intraventricular stasis and could be beneficial for decreasing thrombogenicity.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...