Related Experiment Video
Updated: Jun 30, 2026

GC-based Detection of Aldononitrile Acetate Derivatized Glucosamine and Muramic Acid for Microbial Residue Determination in Soil
Published on: May 19, 2012
A Novel Validated GC-MS/MS Method for the Estimation of N-Nitroso Dimethyl Amine and N-Nitroso Diethyl Amine in
Narayanareddy Papadasu1,2, Ramakrishnareddy Kotanka2, Raghuram Pannala1
1Chemeca Drugs Private Limited, Visakhapatnam 531019, India.
Abstract:
A novel method has been developed for the estimation of N-Nitroso dimethyl amine impurities (NDMA) and N-Nitroso diethyl amine (NDEA) in Zidovudine by using Gas chromatograph Triple Quadrupole Mass with Liquid autosampler (GC-MS/MS) and the method is validated as per International Conference on Harmonization recommendations. Sample analysis was executed for Zidovudine by developed method. Both NDMA and NDEA were detected in below quantitation limit for the Zidovudine batches. Efficient chromatographic separation was achieved on a DB-WAX 30 m length × 0.25 mm internal diameter, 0.5-μm film thickness, Triple quad-8040 GC-MS/MS. Quantification was carried out at Triple quad electron ionization source was at a column flow of 1.5 mL/min at a column oven temperature 50°C. The precision was in the range of 0.9-2.5% for NDMA and 0.8-2.3% for NDEA, and regression analysis shows as r value (correlation coefficient) of is >0.99. This method is capable to detect the NDMA and NDEA impurities in Zidovudine at a level of 0.006 ppm for limit of detection and 0.018 ppm for limit of quantitation with respect to test concentration of 66.66 mg/mL.
More Related Videos
07:38A General Method for Detecting Nitrosamide Formation in the In Vitro Metabolism of Nitrosamines by Cytochrome P450s
Published on: September 25, 2017
12:15Quantification of three DNA Lesions by Mass Spectrometry and Assessment of Their Levels in Tissues of Mice Exposed to Ambient Fine Particulate Matter