Setup time analysis for stereotactic body radiotherapy in O-ring linear accelerator without rotational correction
Biplab Sarkar1,2, Anirudh Pradhan3
1Department of Radiation Oncology, Apollo Multispeciality Hospitals, Kolkata, India. biplabphy@gmail.com.
Abstract:
This study analyse setup time (ST) and frequency of on-board imaging for stereotactic abdomen (liver, stomach), lung, and spine radiotherapy in the absence of automatic rotational correction. Total 53 stereotactic body radiotherapy (SBRT) patients, 28 of abdomen, 19 lung, and 6 spine treated for 230 sessions in O-ring gantry accelerator were evaluated for ST analysis. The mean setup time for all patients, abdomen, lung, and spine cases were 7.7 ± 7.4 min, 9.2 ± 9.2 min, 6.3 ± 4.1 min, and 5.5 ± 3.3 min, respectively. Median number CBCT was 2. 96% of cases had a CBCT between 1 and 3, and 9 (4%) had ≥ 4 CBCTs. Overall, 38.1%, 35.5%, 22.1%, 2.2%, and 2.2% of setup time fall into window of 0-5 min, 5-10 min, 10-20 min, 20-30 min, and > 30 min. Most difficult challenge is to negotiate with unknown rotational errors. It will be easy to dealt with them without automatic rotational correction if values are known.
Related Concept Videos
Rotation with Constant Angular Acceleration - I
Using our intuition, we can begin to see how rotational quantities such as angular displacement, angular velocity, angular acceleration, and time are related to one another. For example, if a flywheel...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Relative Motion Analysis using Rotating Axes - Acceleration
Time differentiation is...