Related Experiment Video
Updated: Jun 29, 2025

07:57
Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
13.4K
A Comparative Study of Shade-Matching Reproducibility Using an Intraoral Scanner and a Spectrophotometer
Franciele Floriani1, Carlos A Jurado2, Salahaldeen Abuhammoud1
1Department of Prosthodontics, The University of Iowa College of Dentistry and Dental Clinics, Iowa City, IA 52242, USA.
Dentistry Journal
|March 27, 2024
Summary
Digital tooth color-matching devices, including intraoral scanners and spectrophotometers, show high repeatability at baseline. However, their accuracy for shade determination decreases after aging, especially for darker resin composite shades.
Area of Science:
- Dental Materials Science
- Digital Dentistry
- Color Science
Background:
- Assessing the repeatability of digital tooth color-matching methods is crucial for accurate dental restorations.
- Two primary digital methods, intraoral scanners and spectrophotometers, were evaluated for shade determination.
- The study focused on resin composite restorations and acrylic teeth in light and dark shades, both before and after an aging process.
Purpose of the Study:
- To compare the repeatability and accuracy of two digital tooth color-matching devices.
- To evaluate the impact of an aging process on shade determination accuracy.
- To assess performance across different shades (light and dark) of resin composites and acrylic teeth.
Main Methods:
- Forty acrylic teeth (shades A1, A3) were restored with resin composite (shades A1, A3) in standardized Class V cavities.
- Aging was simulated using 120 hours of ultraviolet light exposure.
- Shade measurements were taken five times at baseline and after aging using a Trios 3Shape intraoral scanner and a Vita Easyshade Compact spectrophotometer under D65 illumination.
Main Results:
- Both devices demonstrated 100% repeatability for artificial teeth at baseline.
- After aging, trueness for artificial teeth (shade A1) was 80% for both devices.
- For Class V restorations, trueness varied: intraoral scanner (A1: 80%, A3: 60%) and spectrophotometer (A1: 60%, A3: 60%) at baseline.
Conclusions:
- Intraoral scanners and spectrophotometers are reliable for initial shade selection.
- The accuracy of these digital shade-matching devices diminishes following aging.
- Darker shades are particularly susceptible to reduced accuracy after aging.