Technical note: A comparison of physician doses in C-Arm and CT fluoroscopy procedures
Jonathan Troville1, Emily Knott2, Carlos Alberto Reynoso-Mejia1
1Departments of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Purpose:
We address the misconception that the typical physician dose is higher for CT fluoroscopy (CTF) procedures compared to C-Arm procedures.
Methods:
We compare physician scatter doses using two methods: a literature review of reported doses and a model based on a modified form of the dose area product (DAP). We define this modified form of DAP, "cumulative absorbed DAP," as the product of the area of the x-ray beam striking the patient, the dose rate per unit area, and the exposure time.
Results:
The patient entrance dose rate for C-Arm fluoroscopy (0.2 mGy/s) was found to be 15 times lower than for CT fluoroscopy (3 mGy/s). A typical beam entrance area for C-Arm fluoroscopy reported in the literature was found to be 10.6 × 10.6 cm (112 cm2), whereas for CTF was 0.75 × 32 cm (24 cm2). The absorbed DAP rate for C-Arm fluoroscopy (22 mGy*cm2/s) was found to be 3.3 times lower than for CTF (72 mGy*cm2/s). The mean fluoroscopy time for C-Arm procedures (710 s) was found to be 21 times higher than for CT fluoroscopy procedures (23 s). The cumulative absorbed DAP for C-Arm procedures was found to be 9.4 times higher when compared to CT procedures (1.59 mGy*m2 vs. 0.17 mGy*m2).
Conclusions:
The higher fluoroscopy time in C-Arm procedures leads to a much lower cumulative DAP (i.e., physician scatter dose) in CTF procedures. This result can inform interventional physicians deciding on whether to perform inter-procedural imaging inside the room as opposed to retreating from the room.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography