Related Experiment Video
Updated: May 13, 2026

In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System
Sabrina L Lanker1, Christopher J Pino2, H David Humes1
1Innovative BioTherapies Inc, Ann Arbor, Michigan, USA.
Introduction:
The manual single lumen alternating micro-batch hemodiafiltration (mSLAMB) system is a closed-loop dialysis system designed to provide kidney support in emergency situations (e.g., fluid overload, hyperkalemia, acidemia). If done repeatedly in small batches and at high flow rates, this system was found to achieve clearance levels comparable to traditional renal replacement therapy.
Methods:
Using a porcine model, uremic toxins and exogenous fluorescent tracer concentrations were successfully lowered within just 1 h of treatment.
Results:
With a maximal dialysate flow, mSLAMB can achieve decreases in serum potassium concentration of >0.5 mmol/L/h. With the mSLAMB hemodiafiltration system, micro-batch processing was also successful in removing up to 250 mL of ultrafiltrate in 8 cycles.
Conclusion:
This process could create a better fluid balance allowing for administering therapeutic fluids such as sodium bicarbonate in the clinic. Electrolyte imbalance and volume overload remain severe life-threatening emergencies in low resource settings; therefore, mSLAMB should be explored further due to its modest vascular access requirements, low cost, and ability to be performed without electricity or batteries.
More Related Videos
10:19Three-Dimensionally Printed Microfluidic Cross-flow System for Ultrafiltration/Nanofiltration Membrane Performance Testing
Published on: February 13, 2016
06:26Author Spotlight: Advancing Eye Physiology Research via a Multi-Channel Flow Culture for Optimal Tissue Maintenance and Real-Time Assessment
Published on: July 14, 2023
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications