Comprehensive radiation shield minimizes operator radiation exposure in coronary and structural heart procedures
David G Rizik1, Kevin P Gosselin2, Robert F Burke1
1Honor Health, Scottsdale, AZ, United States of America.
Objectives:
This study evaluated the efficacy of a novel comprehensive shield designed to minimize radiation exposure (RE) to Physicians performing coronary and structural heart procedures.
Background:
The Protego™ radiation shielding system (Image Diagnostics Inc., Fitchburg, Ma) is designed to provide comprehensive protection from RE and has been State certified sufficient to allow operators to perform procedures without orthopedically burdensome lead aprons.
Methods:
This single center two-group cohort study assessed the efficacy of this shield in a large number of cardiac procedures (coronary and structural), comparing operator RE compared to standard protection methods (personal lead apparel and "drop down" shield).
Results:
The Protego™ system reduced operator RE by 99 % compared to Standard Protection. RE was significantly lower at both "Head" level by thyroid median dose 0.0 (0.0, 0,0) vs 5.7 (2.9, 8.2) μSv (p < 0.001), as well as waist dose 0.0 (0.0, 0.0) vs 10.0 (5.0, 16.6) μSv (p < 0.001). "Zero" Total RE was documented by Raysafe™ in 64 % (n = 32) of TAVR cases and 73.2 % (n = 183) of the coronary cases utilizing Protego™. In contrast, standard protection did not achieve "Zero" exposure in a single case. These dramatic differences in RE were achieved despite higher fluoroscopy times in the Protego™ arm (11.9 ± 8.6 vs 14.3 ± 12.5 min, p = 0.015). Per case procedural exposure measured by Dose Area Product was higher in the Protego™ group compared to standard protection (115.4 ± 139.2 vs 74.9 ± 69.3, p < 0.001).
Conclusion:
The Protego™ shield provides total body RE protection for operators performing both coronary and structural heart procedures. This shield allows procedural performance without the need for personal lead aprons and has potential to reduce catheterization laboratory occupational health hazards.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...