Validation of a new non-hyperemic physiological index: the constant-resistance ratio (cRR)
Chenguang Li1, Jing Wu2, Jiayan Lin3
1Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai, China.
Objectives:
The instantaneous wave-free ratio (iwFR) has limited availability. A new resting index called the constant-resistance ratio (cRR), which dynamically identifies cardiac intervals with constant and minimum resistance, has been developed; however, its diagnostic performance is unknown. The aim of this study was to validate the cRR by retrospectively calculating the cRR values from raw pressure waveforms of 2 publicly available datasets and compare them with those of the iwFR.
Methods:
Waveform data from the CONTRAST and VERIFY 2 studies were used. The primary endpoint was Bland-Altman bias between cRR and iwFR. Secondary endpoints included diagnostic agreement, correlation, receiver operating characteristic (ROC) analysis, and success rates of cRR and iwFR.
Results:
Among the 1036 waveforms, 871 were successful in determining paired cRR and iwFR values, while cRR was 6% more successful than iwFR (P less than .0001). The mean bias between cRR and iwFR was 0.003, with 95% limits of agreement [-0.021,0.028]. These 2 indices were highly correlated (r = 0.991; P less than .0001). Using an iwFR of 0.89 or less as the reference standard, the optimal cRR cutoff was 0.89, with an area under the ROC curve of 0.991 (P less than .001) and a diagnostic accuracy of 96.9% (95% CI [96%, 98%]).
Conclusions:
The cRR, a new resting index for identifying dynamic cardiac intervals with constant and minimum resistance, demonstrated high numerical agreement, diagnostic consistency, and a higher success rate than the iwFR based on the 2 publicly available datasets.
More Related Videos
Related Concept Videos
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Pre-Procedural Guidelines for Assessing Blood Pressure
Blood Pressure
Measurement of Blood Pressure
Relative Risk
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...