Numerical simulation of the distal stent graft-induced new entry after TEVAR
Meixuan Li1, Tao Ma2, Yunhan Cai1
1Department of Aeronautics and Astronautics, Institute of Biomechanics, Fudan University, Shanghai, China.
Abstract:
The study aimed to investigate the mechanical factors for distal stent graft-induced new entry (dSINE) in aortic dissection patients and discussed these factors in conjunction with aortic morphology. Two patients (one dSINE and one non-dSINE), with the same age, gender, and type of implanted stent, were selected, then aortic morphological parameters were calculated. In addition, the stent material parameters used by the patients were also fitted. Simulations were performed based on the patient's aortic model and the stent graft used. The true lumen segment at the distal stent graft was designated as the "dSINE risk zone," and mechanical parameters (maximum principal strain, maximum principal stress) were computed. When approaching the area with higher mechanical parameters in the dSINE risk zone, dSINE patient exhibited higher values and growth rates in mechanical parameters compared to non-dSINE patient. Furthermore, dSINE patient also presented larger aortic taper ratio, stent oversizing ratio, and expansion mismatch ratio of the distal true lumen (EMRDTR). The larger mechanical parameters and growth rates in dSINE patient corresponded to a greater aortic taper ratio, stent oversizing ratio, and EMRDTR. The failure of dSINE prediction by the stent tortuosity index indicated that mechanical parameters were the fundamental reasons for dSINE development.
More Related Videos
05:04Implantation of Human-Sized Coronary Stents into Rat Abdominal Aorta Using a Trans-Femoral Access
Published on: November 19, 2020
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016