Study of a plastic scintillating plate-based quality assurance system for pencil beam scanning proton beams
Jaehyeon Seo1, Kwangzoo Chung2, Youngyih Han2,2
1Department of Bio-Convergence Engineering, Korea University, Seoul, Republic of Korea.
Introduction:
The purpose of this study was to evaluate a plastic scintillating plate-based beam monitoring system to perform quality assurance (QA) measurements in pencil beam scanning proton beam.
Methods:
Single spots and scanned fields were measured with the high-resolution dosimetry system, consisting of a plastic scintillation plate coupled to a camera in a dark box at the isocenter. The measurements were taken at 110-190 MeV beam energies with 30° gantry angle intervals at each energy. Spot positions were determined using the plastic scintillating plate-based dosimetry system at the isocenter for 70-230 MeV beam energies with 30° gantry angle intervals. The effect of gantry angle on dose distribution was also assessed by determining the scanning pattern for daily QA and 25 fields treated with intensity-modulated proton therapy.
Results:
Spot size, field flatness, and field symmetry of plastic scintillating plate-based dosimetry system were consistent with EBT3 at all investigated energies and angles. In all investigated energies and angles, the spot size measured was ±10% of the average size of each energy, the spot position measured was within ±2 mm, field flatness was within ±2%, and field symmetry was within ±1%. The mean gamma passing rates with the 3%/3 mm gamma criterion of the scanning pattern and 25 fields were 99.2% and 99.8%, respectively.
Conclusions:
This system can be effective for QA determinations of spot size, spot position, field flatness, and field symmetry over 360° of gantry rotation in a time- and cost-effective manner, with spatial resolution comparable to that of EBT3 film.