Investigation on the Validity of in vitro Sun Protection Factor and Ultraviolet-A Protection Factor Evaluation Method
Miyuki Fujishiro1,2, Shoichi Yahagi3,2, Akihiro Kuroda4,2
1New Business Promotion Division, Nikoderm Research Inc.
The ISO24443 method for sunscreen testing shows reliable in vitro Sun Protection Factor (SPF) and Ultraviolet-A Protection Factor (UVA-PF) results. Spectral irradiance sources, not spectrophotometers, significantly impact SPF and UVA-PF outcomes.
Area of Science:
- Dermatology
- Photoprotection
- Analytical Chemistry
Background:
- The ISO24443 standard provides a method for in vitro determination of sunscreen protection.
- Ensuring the reliability and validity of this method is crucial for accurate product labeling and consumer safety.
- Previous studies have highlighted potential variability in in vitro sunscreen testing.
Purpose of the Study:
- To analyze factors influencing in vitro Sun Protection Factor (SPF) and Ultraviolet-A Protection Factor (UVA-PF) measurements.
- To verify the validity and reliability of the ISO24443 evaluation method for sunscreens.
- To identify sources of variability in in vitro SPF and UVA-PF determinations.
Main Methods:
- In vitro measurements of UV absorbance using various spectrophotometers.
- Comparison of SPF and UVA-PF results obtained with different spectral irradiance sources, including European Cosmetic and Perfumery Association (Colipa) 1994 and UV-solar simulated radiation (UV-SSR).
- Analysis of the impact of spectrophotometer differences and spectral irradiance on calculated SPF and UVA-PF values.
Main Results:
- UV absorbance measurements showed minimal differences across different spectrophotometers, even with significant absorbance variations above 2.
- In vitro SPF and UVA-PF values exhibited considerable differences when using Colipa 1994 or UV-SSR spectral irradiance.
- The necessity for re-evaluating the appropriateness of the coefficient of adjustment for in vitro UVA-PF determination was indicated.
Conclusions:
- The ISO24443 method demonstrates good reliability concerning spectrophotometer variability for in vitro SPF and UVA-PF.
- Spectral irradiance sources are a critical factor influencing the accuracy of in vitro SPF and UVA-PF results.
- Further investigation into adjustment coefficients is recommended for precise in vitro UVA-PF assessment.