Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Remote and low-cost intraocular pressure monitoring by deep learning of speckle patterns
Zeev Kalyuzhner1, Sergey Agdarov1, Yevgeny Beiderman2
1Bar-Ilan University, Faculty of Engineering and the Nanotechnology Center, Ramat-Gan, Israel.
Significance:
Glaucoma, a leading cause of global blindness, disproportionately affects low-income regions due to expensive diagnostic methods. Affordable intraocular pressure (IOP) measurement is crucial for early detection, especially in low- and middle-income countries.
Aim:
We developed a remote photonic IOP biomonitoring method by deep learning of the speckle patterns reflected from an eye sclera stimulated by a sound source. We aimed to achieve precise IOP measurements.
Approach:
IOP was artificially raised in 24 pig eyeballs, considered similar to human eyes, to apply our biomonitoring method. By deep learning of the speckle pattern videos, we analyzed the data for accurate IOP determination.
Results:
Our method demonstrated the possibility of high-precision IOP measurements. Deep learning effectively analyzed the speckle patterns, enabling accurate IOP determination, with the potential for global use.
Conclusions:
The novel, affordable, and accurate remote photonic IOP biomonitoring method for glaucoma diagnosis, tested on pig eyes, shows promising results. Leveraging deep learning and speckle pattern analysis, together with the development of a prototype for human eyes testing, could enhance diagnosis and management, particularly in resource-constrained settings worldwide.