Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
In vitro evaluation of a new viscoelastometry-based point-of-care analyzer
Vanina Corominas1,2, Thomas Chiniard2,3, Pierre Pasquier2,3
1French Military Blood Institute, Clamart, France.
Introduction:
The VCM is a point-of-care analyzer using a new viscoelastometry technique for rapid assessment of hemostasis on fresh whole blood. Its characteristics would make it suitable for use in austere environments. The purpose of this study was to evaluate the VCM in terms of repeatability, reproducibility and interanalyzer correlation, reference values in our population, correlation with standard coagulation assays and platelet count, correlation with the TEG5000 analyzer and resistance to stress conditions mimicking an austere environment.
Methods:
Repeatability, reproducibility, and interanalyzer correlation were performed on quality control samples (n = 10). Reference values were determined from blood donor samples (n = 60). Correlations with standard biological assays were assessed from ICU patients (n = 30) and blood donors (n = 60) samples. Correlation with the TEG5000 was assessed from blood donor samples. Evaluation of vibration resistance was performed on blood donor (n = 5) and quality control (n = 5) samples.
Results:
The CVs for repeatability and reproducibility ranged from 0% to 11%. Interanalyzer correlation found correlation coefficients (r2) ranging from 0.927 to 0.997. Our reference values were consistent with those provided by the manufacturer. No robust correlation was found with conventional coagulation tests. The correlation with the TEG5000 was excellent with r2 ranging from 0.75 to 0.92. Resistance to stress conditions was excellent.
Conclusion:
The VCM analyzer is a reliable, easy-to-use instrument that correlates well with the TEG5000. Despite some logistical constraints, the results suggest that it can be used in austere environments. Further studies are required before its implementation.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
In Vitro Drug Release Testing: Overview, Development and Validation
Automated Microbial Diagnostics