Flexible low-profile external ventricular drain catheter for real-time brain monitoring
Hoon Yi1, Hodam Kim1, Ka Ram Kim1
1George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USA; IEN Center for Wearable Intelligent Systems and Healthcare Institute for Electronics and Nanotechnology, Georgia Institute of Technology, Atlanta, GA, 30332, USA.
Abstract:
External ventricular drainage is one of the most common neurosurgical procedures in the world for acute hydrocephalus, which must be performed carefully by a neurosurgeon. Although various neuromonitoring external ventricular drain (EVD) catheters have been utilized, they still suffer from rigidity and bulkiness to mitigate post-EVD placement trauma. Here, we introduce a flexible and low-profile smart EVD catheter using a class of technologies with sensitive electrical materials, seamless integration, and flexible mechanics, which serves as a highly soft and minimally invasive device to monitor electrical brain signals. This device reliably captures biopotentials in real time while exhibiting remarkable flexibility and reliability. The seamless integration of its sensory system promises a minimally invasive EVD placement on brain tissue. This work validates the device's distinct characteristics and performances through in vitro experiments and computational analysis. Collectively, this device's exceptional patient- and user-friendly attributes highlight its potential as one of the most practical EVD catheters.
More Related Videos
14:58Comprehensive Endovascular and Open Surgical Management of Cerebral Arteriovenous Malformations
Published on: October 20, 2017
09:16Use of a Wireless Video-EEG System to Monitor Epileptiform Discharges Following Lateral Fluid-Percussion Induced Traumatic Brain Injury
Published on: June 21, 2019