Comparative effects of variable versus conventional volume-controlled one-lung ventilation on gas exchange and
Jakob Wittenstein1, Robert Huhle1, Anne-Kathrin Mutschke1
1Department of Anesthesiology and Intensive Care Medicine, Pulmonary Engineering Group, University Hospital Carl Gustav Carus Dresden, TUD Dresden University of Technology, Dresden, Germany.
Background:
Mechanical ventilation with variable tidal volumes (V-VCV) has the potential to improve lung function during general anesthesia. We tested the hypothesis that V-VCV compared to conventional volume-controlled ventilation (C-VCV) would improve intraoperative arterial oxygenation and respiratory system mechanics in patients undergoing thoracic surgery under one-lung ventilation (OLV).
Methods:
Patients were randomized to V-VCV (n = 39) or C-VCV (n = 39). During OLV tidal volume of 5 mL/kg predicted body weight (PBW) was used. Both groups were ventilated with a positive end-expiratory pressure (PEEP) of 5 cm H2O, inspiration to expiration ratio (I:E) of 1:1 (during OLV) and 1:2 during two-lung ventilation, the respiratory rate (RR) titrated to arterial pH, inspiratory peak-pressure ≤ 40 cm H2O and an inspiratory oxygen fraction of 1.0.
Results:
Seventy-five out of 78 Patients completed the trial and were analyzed (dropouts were excluded). The partial pressure of arterial oxygen (PaO2) 20 min after the start of OLV did not differ among groups (V-VCV: 25.8 ± 14.6 kPa vs C-VCV: 27.2 ± 15.3 kPa; mean difference [95% CI]: 1.3 [-8.2, 5.5], P = 0.700). Furthermore, intraoperative gas exchange, intraoperative adverse events, need for rescue maneuvers due to desaturation and hypercapnia, incidence of postoperative pulmonary and extra-pulmonary complications, and hospital free days at day 30 after surgery did not differ between groups.
Conclusions:
In thoracic surgery patients under OLV, V-VCV did not improve oxygenation or respiratory system mechanics compared to C-VCV. Ethical Committee: EK 420092019.
Trial Registration:
at the German Clinical Trials Register: DRKS00022202 (16.06.2020).
More Related Videos
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Respiratory Volumes and Capacities
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...