Novel self-expandable stent-based endobiliary radiofrequency ablation for unresectable malignant biliary obstruction
Yubeen Park1, Dong-Sung Won2, Jin Su An3
1Biomedical Engineering Research Center, Asan Institute for Life Sciences, Asan Medical Center, Seoul, Republic of Korea; Department of Gastroenterology, Asan Medical Center, Seoul, Republic of Korea.
Background And Aims:
Endobiliary radiofrequency ablation (RFA) is an emerging endoscopic palliative adjunctive therapy used for the local treatment of unresectable malignant biliary obstruction (MBO). However, irregular ablation ranges caused by insufficient electrode-to-bile duct contact pose a significant obstacle. We investigated the feasibility of a self-expandable stent (SES)-based electrode with a customized RFA generator in the porcine liver and common bile duct (CBD).
Methods:
An SES-RFA system with polarity switching was developed to perform endobiliary RFA. The ablation ranges of 20 ablation protocols were evaluated to validate the feasibility of the newly developed RFA system in the porcine liver. Nine of 20 ablation protocols were selected for evaluation in the porcine CBD with cholangiography, endoscopy, and histologic and immunohistochemical analysis.
Results:
The SES-RFA system with polarity switching was successfully constructed and demonstrated high accuracy and reproducibility. The ablation area was clearly identified between the 2 SESs. The ablation ranges and degree of mucosal damage, including terminal deoxynucleotidyl transferase-mediated dUTP nick and labeling-positive and heat shock protein 70-positive depositions, increased proportionally with ablation protocols in the porcine liver and CBD (all P < .05). Ablation length and depth linearly increased with ablation protocols from 8.74 ± .25 to 31.25 ± .67 mm and 1.61 ± .09 to 11.94 ± .44 mm, respectively.
Conclusions:
The SES-RFA system with polarity switching between electrodes provided an even circumferential area of ablation and enhanced ablation depth between the electrodes. This novel endobiliary RFA system is a promising modality for local ablation in patients with unresectable MBO.